Measurement wavelength
|
380nm - 780nm
|
Measurement method
|
Comparison measurement using reference specimen
|
Sample N.A.*
|
0.12 (using 10x objective lens)
0.24 (using 20x objective lens)
* it differs from objective lens' N.A.
|
Sample W.D.
|
10.1 mm (using 10x objective lens)
3.1 mm (using 20x objective lens)
|
Sample curvature radius
|
-1R - ∞, +1R - ∞
|
Sample measurement range
|
Approximately ø60 µm (using 10x objective lens)
Approximately ø30 µm (using 20x objective lens)
|
Measurement repeatability (2σ)
|
±0.1 % and less (during 380 nm - 410 nm measurement)
±0.01 % and less (during 410 nm - 700 nm measurement)
|
Display resolution
|
1 nm
|
Measurement time
|
A few secs-several secs (it differs from sampling time)
|
Light source
|
Halogen lamp 12 V 100W
|
Z-direction movement range of the stage
|
85mm
|
PC interface
|
USB
|
Weight
|
Unit Approximately 20 kg
Power source for light source: Approximately 3 kg
Control box: Approximately 8 kg
|
External dimensions
|
Unit: 300 (W) x 550 (D) x 570 (H) mm
Power source for light source: 150 (w) x 250 (D) x 140 (H) mm
Control box: 220 (W) x 250 (D) x 140 (H) mm
|
Power Source
|
Power source for light source: 100 V (2.8A)/220 V AC
Control box: 100 V (0.2A)/220V AC
|
Environment for use
|
Installed at the horizontal plane with no vibration
Temperature: 23 ±5°C
Humidity: 60% or less
An environment is not subject to air fluctuation
|